íÌÈ ãæÇÝÞÉ  ÓÚÇÏÉ ÇáÏßÊæÑ       /ÚãíÏ ÇáßáíÉ    æÓÚÇÏÉ ÇáÏß澄 /ÑÆíÓ ÇáÞÓã  Úáì ØáÈÇÊ       ÇáÊÍáíá
  ãæÞÚ ÎÇÕ.........ãÚÇãá ÞÓã-ÇáËÑæÉ ÇáãÚÏäíÉ æÇáÕÎæÑ

home

visitors      

Mineral Resources and Petrology lab   

Search this site or the web powered by FreeFind

Site search Web search

A Guide to

Scanning Microscope

Observation

Preface Preface Preface Preface Preface

Today, the scanning electron microscope (hereinafter ab-breviated

to SEM) is utilized not only in medical science and

biology, but also in diverse fields such as materials develop-ment,

metallic materials, ceramics, and semiconductors.

This instrument is getting easier to use with the progress

of electronics and introduction of new techniques. Anybody

can now take micrographs after short-time training in its op-erational

procedure. However, when one has begun to use

the instrument, he cannot always take satisfactory photos.

When the photo is not sharp enough, or when necessary

information cannot be obtained, it is necessary to think what

causes it.

To help make a correct judgment in such a case, the first

edition of “A Guide to Scanning Microscope Observation” was

published and it has since been used by many people. To-day,

when several years have passed since the publication

of the first edition, some parts of the edition need amend-ment

with instrumental improvements. This is the reason

why we bring this revised edition to you.

We included in this book as many application examples as

possible so that they can be used as criteria for judging what

causes unsatisfactory image factors (hereinafter referred to

as image disturbances). Although this edition does not de-scribe

all about image disturbances, it carries application

photos to allow you to consider their causes. It is also im-portant

to correctly select the optimum observation condi-tions

fcr various specimens. For instance, this book carries

matters which are considered to be useful for using the in-strument,

such as the accelerating voltage, probe current

and working distance (hereinafter abbreviated to WD).

We shall be pleased if this publication is of help to people

who are now using or going to use SEMS..4

Lack of sharpness

Low image quality

Noises

Image distortion and deformation

the basis of the data available at hand, for cases such as the

following:

1) The mutual interaction between the specimen and

the electron beam involves a problem.

2) Selection of observation conditions and specimen

preparation involve a problem.

3) The instrument itself involves a problem.

Types of Image Disturbances

Image disturbances, though diverse in types, can be classified

by the following expressions:

1) Images lacking sharpness and contrast

2) Unstable images

3) Generally poor-quality images

4) Noisy images

5) Images showing jagged edges

6) Unusual-contrast images

7) Distorted or deformed images.

The above-listed image disturbances, besides being attributed

to defects in the instrument itself, are occasionally caused by the

operator’s lack of experience, improper specimen preparation and

external influences such as the installation room conditions.Table

1 shows various image disturbances and their causes.

In this booklet, how image disturbances appear is studied on

2-1. Influence of accelerating voltage

on image quality

When theoretically considering the electron probe diameter

alone, the higher the accelerating voltage, the smaller is the elec-tron

probe. However, there are some unnegligible demerits in in-creasing

the accelerating voltage. They are mainly as follows:

1) Lack of detailed structures of specimen surfaces.

2) Remarkable edge effect.

3) Higher possibility of charge-up.

4) Higher possibility of specimen damage.

Image Changes Caused by Interactions Between Electron Probe

and Specimen

In SEM, finer surface structure images can generally be ob-tained

with lower accelerating voltages. At higher accelerating

voltages, the beam penetration and diffusion area become larger,

resulting in unnecessary signals (e.g., backscattered electrons)

being generated from within the specimen. And these signals

reduce the image contrast and veils fine surface structures. It is

especially desirable to use low accelerating voltage for observa-tion

of low-concentration substances.

2-2. Probe current, probe diameter,

and image quality

In the SEM, the smaller the electron probe diameter on the

specimen, the higher the magnification and resolution. However,

the image smoothness, namely, the S/N ratio depends on the

probe current. The probe current and the probe diameter are in

the relation shown in Fig. 8. Namely, as the probe diameter is

reduced, the probe current is reduced.

It is therefore necessary to select a probe current suited for the

magnification and observation conditions (accelerating voltage,

specimen tilt, etc.) and the specimen.

2-3. Influence of edge effect on image

quality

Among the contrast factors for secondary electrons, the tilt ef-fect

and edge effect are both due to the specimen surface mor-phology.

Secondary electron emission from the specimen sur-face

depends largely on the probe’s incident angle on the speci-men

surface, and the higher the angle, the larger emission is

caused. The objects of the SEM generally have uneven surfaces.

There are many slants all over them, which contribute most to

the contrast of secondary electron images.

On the other hand, large quantities of secondary electrons are

generated from the protrusions and the circumferences of ob-jects

on the specimen surface, causing them to appear brighter

than even portions.

The degree of the edge effect depends on the accelerating volt-age.

Namely, the lower the accelerating voltage, the smaller the

penetration depth of incident electrons into the specimen. This

reduces bright edge portions, thus resulting in the microstruc-tures

present in them being seen more clearly.

Normally, secondary electron images contain some

backscattered electron signals. Therefore, if the tilt direction of

the specimen surface and the position of the secondary electron

detector are geometrically in agreement with each other, more

backscattered electrons from the tilted portions are mixed, caus-ing

them to be seen more brightly due to synergism. (a) 5 kV x720 Tilt Angle: 50°

(b) 5 kV x720 Tilt Angle: 50°

(a) Protrusion (b) Edge (c) Circumference.10

2-4. Use of specimen tilt

Specimen tilt is aimed at:

1) Improving the quality of secondary electron images

2) Obtaining infromation different form that obtained when

the specimen is not tilted, that is, observing topographic

features and observing specimen sides.

3) Obtaining stereo micrographs.

a) Dependence of image quality on tilt angle

Fig. 13 shows a photo taken at a tilt angle of 0° (a) and a photo

taken at 45° (b). Their comparison shows that the latter is of smooth

quality and stereoscopic as compared with the former. When the

specimen is tilted, however lengths observed are different from

their actual values. When measuring pattern widths, etc., there-fore,

it is necessary to measure without specimen tilting or to cor-rect

values obtained form a tilted state.

b) Stereo micrographs

With SEM images it is sometimes difficult to correctly judge their

topographical features. In such a case observation of stereo SEM

images makes it easy to understand the structure of the speci-men.

Besides, stereo observation allows unexpected information

to be obtained even from specimens of simple structure.

In stereo observation, after a field of interest is photographed,

the same field is photographed again with the specimen tilted

from 5º to 15º. Viewing these two photos using stereo glasses

with the tilting axis held vertically provides a stereo image.

(a) Tilt angle: 0°

(b) Tilt angle: 45°

c) Detector position and specimen direction

The amount of secondary electrons produced when the speci-men

is illuminated with an electron beam, depends on the angle

of incidence theoretically. However, there arises a difference in

the image brightness depending on whether the tilted side of the

specimen is directed to the secondary electron detector or th

eopposite side.

With a long specimen, for example, the brightness differes be-tween

the side facing the detector and the opposite side.

In such a case, directing the longitudinal axis of the speciment

to the detector makes the brightness uniform.

(a) Specimen directed as at 1

(c) Specimen directed as at 3

(b) Specimen directed as at 2

2-5. Use of backscattered electron

signals

Although secondary electron images are obtained most fre-quently

with the SEM, backscattered electron images also pro-vide

important information.

Backscattered electrons vary in their amount and direction with

the composition, surface topography, crystallinity and magnetism

of the specimen. The contrast of a backscattered electron image

depends on (1) the backscattered electron generation rate that

depends on the mean atomic number of the specimen, (2) angle

dependence of backscattered electrons at the specimen surface,

and (3) the change in the backscattered electron intensity when

the electron probes incident angle upon a crystalline specimen

is changed.

The backscattered electron image contains two types of infor-mation:

one on specimen composition and the other on speci-

men topography. To separate these two types of information, a

paired semiconductor detector is provided symmetrically with

respect to the optical axis. Addition of them gives a composition

image while subtraction gives a topography image. And with com-position

images of crystalline specimens, the difference in crystal

orientation can be obtained as the so-called “channeling con-trast,”

by utilizing the advantage that the backscattered electron

intensity changes largely before and after Bragg’s condition.

The generation region of backscattered electrons is larger than

that of secondary electrons, namely, several tens of nm. There-fore,

backscattered electrons give poorer spacial resolution than

secondary electrons. But because they have a larger energy than

secondary electrons, they are less influenced by charge-up and

specimen contamination.

2-6. Influence of charge-up on image

quality

a) Charge-up and countermeasure against it

When a nonconductive specimen is directly illuminated with an

electron beam, its electrons with a negative charge collect locally

(specimen charge-up), thus preventing normal emission of sec-ondary

electrons. This charge-up causes some unusual phenom-ena

such as abnormal contrast and image deformation and shift.

Usually, the surface of a nonconductive specimen is coated with

some conductive metal prior to observation. Rough surfaced

specimens must be evenly coated from every direction. Recently,

however, a method has been employed to observe specimens

without coating, in order to know their true surface state.

Generally, the following methods are used to reduce specimen

charge-up.

1) Reducing the probe current

2) Lowering the accelerating voltage

3) Tilting the specimen to find a balanced point between the

amount of incident electrons and the amount of

electrons that go out of the specimen (this point varies

with the specimen).

(a) 1.0 kV x3,200 (a) 4 kV

(b) 1o kV (b) 1.3 kV x3,200

b) Prevention of charge-up by sampling

Biological, cloth, and powder specimens cannot often be pho-tographed

clearly, with some portions looking too bright and some

too dark. This is because those specimens are partly charged

up.

To prevent this, it is necessary to give specimen surfaces uni-form

conductivity as follows:

(1) When fixing the specimen on a specimen stub, apply

conductive paint (carbon paint or the like) to specimen

portions which are hard to coat.

(2) In the case of powder, if its particles are piled on each

other, charge-up easily takes place, causing them to

move during observation. To prevent this, after the

adhesive for fixing the power is dried, blow the piled

particles using a hand blower. Different adhesives need

to be used depending on the size of particles. Mainly,

double-sided adhesive tape, manicure liquid, and

aluminum foil are used. When using double-sided tape,

it is effective to apply conductive paint (carbon paint) to

the four corners.

(a) 15 kV Tilt angle: 0° x720

(b) 5 kV Tilt angle: 45° x1,400

(c) 15 kV Tilt angle: 0° x720

(d) 5 kV Tilt angle: 45° x1,400

Evaporated film

Specimen

Adhesive

Specimen stub

(b) Insufficient adhesive (a) Insufficient adhesive

Fixing of a bulk specimen

Carbon paint

Fixing of fiber

Specimen stub

Adhesive

Particles

Remove excessive particles.16

2-7. Specimen damage by electron

beam

The loss of electron beam energy in the specimen occurs mostly

in the form of heat generation at the irradiated point. The tempera-ture

increase at an irradiated point is dependent on:

1) The electron beam accelerating voltage and dosage.

2) Scanning area.

3) Scanning time.

4) Heat conductivity of the specimen. Polymer materials and

biological specimens, which are gener- ally not resistant

to heat, are easily damaged by the electron beam,

because of their low heat conductivity.

To avoid this damage, the following should be taken into consid-eration:

1) To use low accelerating voltage.

2) To decrease electron beam intensity.

3) To shorten exposure time, even though this reduces

photograph smoothness slightly.

4) To photograph large scanning areas with low

magnifications.

5) To control the thickness of coating metal on the specimen

surface. It is also advisable to adjust beforehand the

astigmatism

and brightness using another field of view and then

photograph the actual field as quickly as possible.

(a) Undamaged specimen

(b) Damaged specimen

When a specimen area is irradiated with an electron

probe for a long tiem at high magnification, it may be

damaaged

2-8. Contamination

When the electron probe is irradiated on a specimen portion for

a long time, its image may lose sharpness and become dark. This

is caused by the residual gas in the vicinity of the specimen being

struck by the electron probe. This phenomenon is called specimen

contamination.

The conceivable residual gases in the specimen chamber, which

cause contamination are:

1) Gas caused from the instrument itself.

2) Gas that specimens bring into the instrument

3) Gas that the specimen itself gives off.

To prevent specimen contamination, special attention must be

paid to the following matters:

1) Use the minimum amount of double-sided adhesive

tape or conductive paint, and completely dry it before

putting the specimen in the instrument.

2) In some cases, contamination can also be reduced by

drying the adhesive with a drier or the like.

3) Use the smallest possible biological specimens.

4) Since some embedding agents and resins give off a large

amount of gas, they need to be selected carefully. Also

since organic gas is given off when the resin surface is

irradiated with an electron probe, irradiate the smallest

possible surface area or coat the surface with a

conductive material.

5 kV x18,000

3. The Influence on Images, of Operational Technique, Specimen

Preparation Technique, External Disturbances, Etc.

3-1. Working distance and objective

aperture

a) Influence of working distance (WD) on images

WD is changeable on many recently available SEM models. Fig.

28 shows what effect is produced on the image when WD is

changed with other conditions kept unchanged.

High Resolution

Greater depth of field

Small

Working Distance

Low resolution

Smaller depth of field

Large

b) Influence of objective aperture diameter on

images

The objective lens (OL) aperture set in the SEM as standard is

of the optimum size selected considering various conditions. SEM

images require not only a fine electron probe, but also a sufficient

amount of signals for forming an image. The aperture cannot be

reduced unnecessarily. The OL aperture must be selected with

consideration given to the effect shown in Fig.

Large

Small

Large current Low resolution

Smaller depth of field

Aperature size

High resolution

Greater depth of field Grainy image

(BEI X-ray analysis)

(a) OL aperature diameter: 600µm WD: 10mm (b) OL aperature diameter: 200µm WD: 10mm

(c) OL aperture diameter: 200µm wd:20mm

(d) OL aperture diameter: 200µm WD: 38mm (e) OL aperture diameter: 100µm WD: 38mm

3-2. Influence of astigmatism

The aberration caused by the machining accuracy and material

of the polepiece is called "astigmatism." This astigmatism can be

removed by adjusting the two knobs, X and Y, of the stigmator.

An image is judged as astigmatism-free if it has no unidirec-tional

defocusing when the objective lens is changed to under or

over-focus at a little high magnification (about x10000).

(A) Images before astigmatism correction

Under focus Just focus

Over focus

Focal line

Minimum circle of confusion

Focal line

(a) Shape changes in electron beam when there is astigmatism.21

Under focus Just focus

Over focus

just focus

(b) Shape changes in electron beam when astigmatism is corrected

The micrographs (f) to (j) show stigmator-corrected images. Al-though

blurring is noticed before and after the just-focus image,

no unidirectional defocusing is seen.

The micrographs (a) to (e) do not provide images as sharp as

the images (f) through (j) due to astigmatism..22

3-3. Optimum contrast and

brightness of micrographs

A good SEM microscope is sharp, noiseless and provides opti-mum

contrast and brightness.

In JEOL SEMs, optimum contrast and brightness are adjusted

automatically or by built in controls. In some cases, however, the

contrast and brightness are adjusted optimumly for the portion of interest only, and not for the image overall.

(1) Excessive contrast

(2) Insufficient brightness (3) Optimum contrast and brightness (4) Excessive brightness

(5) Insufficient contrast

3-4. Exposure time for X-ray image

SEI and BEI images are usually photographed by one scan of

50 to 100 seconds. However, since the signals for an X-ray image

per unit time are small, they are often photographed by long-time

exposure.

If the exposure time is not long enough, X-ray images may lack

some information on element distribution. It is therefore neces-sary

to carefully decide on the exposure time.

Generally,calculating the X-ray count as approx. 2,500 counts/

cm2 gives a good result.

When photographing the distribution of the specific elements in

a certain phase, it is well to fix the electron probe at that portion,

investigate the X-ray count rate (CPS) and the ratio of the phase

to the whole image. The exposure time can than be decided upon.

Exposure time: Time required for photographing

Picture elements The area (cm2) of interest

X-ray count rate: X-ray count rate (CPS) during electron

beam irradiation

How to decide exposure time

Picture elements(cm²)

Exposure time(s) = 2500 (C/cm2>) x ---------------------

X-ray count rate (CPS)

(a) Composition image (COMPO)

(b) X-ray image CA 50 seconds

(c) X-ray image Ca 300 seconds

3-5. Influences of external disturbances

on images

External disturbances such as a stray magnetic field, mechani-cal

vibrations, etc. can cause image distortion, jagged edge lines

and other phenomena. Often disturbances in SEM images are

caused by structural or installation conditions such as:

1) Leakage:

Magnet field from distribution board

High-tension line located too close to the instrument

2) Low floor strength

3) Improper grounding

Installation conditions should be carefully checked in advance

to avoid any problems after SEM installation.

(a) Influenced by external magnetic field.

(b) Influenced by external magnetic field

3-6 Deformation and impurity

precipitation during specimen

preparation process.

In the process of specimen preparation, biological specimens

are apt to become defective becasue of required processes such

as fixing, washing and dehydration. Their defects can be classified

as follows:

1) Specimen deformation

2) Impurity precipitation

3) Impurity coating

To avoid deformation as much as possible, the critical point dry-ing

method is currently used instead of air drying.

Impurity precipitation is a phenomenon wherein the crystal ma-terial

contained in the fixing, washing and dehydrating agents pre-cipitates

during the drying process. Sufficient care must be taken

when handling the above-mentioned chemicals.

(a) Air drying

(b) Critical point drying

3.7 Image distortion and its cause

If correct deflection magnification is lost horizontally or verti-cally,

it results in image distortion. In some cases, latex particles

must be used for checking purposes.

3-8 Coating

Coating used in SEM analysis is aimed mainly at the following:

1) Preventing the charge-up on the specimen surface by

covering it with a conductive material.

2) Increasing secondary electron emission by covering a

specimen of low secondary electron emission with a

metal of high secondary electron yield.

For coating, the vacuum evaporation method and the sputtering

method are generally used. With the improved resolution of the

SEM, coating techniques for high magnification are still under study.

However, various substances are being used i.e., C (for general

analysis), AU, AU-Pd, and Pt, which must be selected depending

on the purpose and magnification. It is necessary to select a coat-ing

suitable for the observation magnification. If the coating is too

thick, its particles become visible while at the same time the struc-tures

of interest are may be obscured.

a) Sputtering device

This device is most widely used for observing specimen surface

morphology. When coating polymer materials that are easily dam-aged

by ion irradiation and electron irradiation, the triode-type

magnetron sputtering device is recommended over the diode sput-tering.

As metals, AU and AU-pd are generally used because they are

easily obtained and generate secondary electrons well. Recently,

however, high-melting metals such as Pt and W have been used

for high magnification observation, because of their high granular-ity.

Generally, the sputtering device is not used for carbon coating

and is not suited for that purpose.

b) Vacuum evaporator

When making surface observation and elementary analysis by

X-ray detection, vacuum evaporation of carbon is carried out most

generally for minimizing an evaporated substance's interference

with detected elements.

Also, when the charge-up on the specimen surface cannot be

prevented simply by coating with AU because of surface rough-ness,

AU coating may be done after carbon evaporation. In either

case, it is necessary to uniformly coat the specimen from all direc-tions

by rotating and tilting it during evaporation

 

4. Disturbances Caused by Instrumental Defects

4-1. Insufficient filament heating

When the filament tip is not at a high enough temperature, due

to insufficient filament heating, a proper cross over point may not

be obtained, making a sharp image impossible to obtain.

If the filament is heated too much, the filament is evaporated

excessively, which results in the generation of whiskers, instability

of the electron probe, or in a shorter filament life. It is important to

set the temperature optimumly.

4-2 Incorrect alignment and centering

of objective aperture

When the column is disassembled for cleaning or when the elec-tron

beam is lost, an operation called "alignment of the column"

allows for the electron beam from the filament to be most effec-tively

collected onto the specimen surface by means of mechani-cal

and electrical alignment.

In this operation, the electron beam from the gun should first be

aligned using a tilting correction knob, then the objective aperture

should be adjusted to make the electron beam pass the objective

lens center.

After replacing or cleaning the objective aperture, it is neces-sary

to adjust the objective aperture position. It is ideal to set the

objective aperture at the center of the objective polepiece. When

the aperture shifts from this position, however, the astigmatism of

the image becomes extremely high, making it impossible to obtain

high-resolution images.

(a) Incorrect alignment

(b) Correct alignment

4-3 10 kV discharge of detector

The secondary electron detector and CRT are supplied with a

high voltage, 10 kV. The poor connection of cables, exfoliation of

the fluorescent paint, and the presence of dust on the metal ring

and fluorescent plane of the detector can all cause discharges.

The images below show images under a 10 kV discharge. The

effect appears to be similar to those of unusual accelerating volt-age

unsatisfactory gun emission and specimen charge-up. The

difference from their effect is that only brightness is changed, with

no defocusing, image cut and image shift observed.

(a) Discharge of a detector

(b) Discharge of CRT

4-4 Burnt and dusty CRT surface

After long usage, the surface of the CRT becomes dusty and

should be cleaned from time to time. The black dots and lines in

the images below were caused by CRT burn-out due to excessive

brightness or by dust.

Technical terms:

SEM:

Scanning electron microscope

Probe current:

The total amount of current to be irradiated on the specimen. It

is controlled between approx. 10-12A to 10-6A. The control is done

by varying the excitation of the SEM's condenser lens. The name

of this knob with this function varies with the type of the instru-ment

having that function, like CONDENSER LENS, PROBE CUR-RENT

and SPOT SIZE.

SEI: Secondary Electron Image:

Secondary electrons are excited secondarily by electrons inci-dent

on the specimen. Since their generation region is as shallow

as approx. 10 nm, the diffusion of incident electrons within the

specimen has little influence on the image, thus allowing the best

resolution to be obtained.

The contrast of secondary electron images depends mainly on

the tilt angle and topoqranhy of the specimen surface.

BEI: Backscattered Electron Image:

After incident electrons are scattered within the specimen some

of them are backscattered while keeping a relatively high energy

and emitted again from the specimen surface. These electrons

are called backscattered electrons. The contrast of the

backscattered electron image depends on the topography of the

specimen surface and on the mean atomic number of the sub-stances

which constitute the specimen. Use of a paired detector

allows separate observation of a topography (TOPO) image and a

composition (COMPO) image.

X-ray image:

A mapping image used to investigate the distribution of a char-acteristic

X-ray image of a specific element.

Under-focus and over-focus:

When objective lens excitation is weakened below the just-fo-cus

position, the focal point position lowers below the specimen

surface. This focus state is called "under-focus." "Over-focus ' op-posite

to "under-focus," is caused when objective lens exciation is

intensified.

CL and OL: Condenser lens and objective lens:

CL controls the probe current and OL focuses the electron probe

on the specimen surface.

WD: Working distance:

The distance from the underside of the objective lens to the

specimen surface.

Probe diamemeter:

Generally, this means the minimum probe diameter that depends

on the accelerating voltage, probe current, and working distance.

 

Concluding Remarks

Today, as the SEM comes into wider usage, quality images can

be obtained even with the minimum necessary operation skills and

knowledge.

This publication is concerned primarily with issues related to

image quality with the understanding that even beginners can un-derstand

matters concerning SEM images and take satisfactory

photos.

Also, the photos in this publication were taken with various types

of instruments.

 

ismail alnatour . All rights reserved.