íÌÈ ãæÇÝÞÉ  ÓÚÇÏÉ ÇáÏß澄 /ÚãíÏ ÇáßáíÉ    æÓÚÇÏÉ ÇáÏß澄 /ÑÆíÓ ÇáÞÓã  Úáì ØáÈÇÊ ÇáÊÍáíá
  ãæÞÚ ÎÇÕ.........ãÚÇãá ÞÓã-ÇáËÑæÉ ÇáãÚÏäíÉ æÇáÕÎæÑ

ismail alnatour                                                          natour@hotmail.com

 spectroscopy lab

 

Search this site or the web powered by FreeFind

Site search Web search

INTRODUCTION TO GRAPHITE FURNACE ATOMIC ABSORPTION

CONSIDERATIONS IN ULTRA TRACE ANALYSIS

Performance Criteria

Differences in graphite furnace performance characteristics require redefinition

of some basic AA performance criteria.Since the magnitude of the graphite fur-nace

signal observed depends on analyte mass rather than concentration,the term

‘‘characteristic mass’’is used as a measure of the sensitivity of the furnace.Char-acteristic

mass is analogous to characteristic concentration for flame AA except

that mass,rather than concentration,is related to absorbance.The characteristic

mass of an analyte is defined as the mass of analyte in picograms required to pro-duce

a peak height signal of 0.0044 absorbance or an integrated peak area signal

of 0.0044 absorbance-seconds (A×s).

Similar to characteristic concentration for flame AA,characteristic mass may be

used as an indicator of instrument optimization.Typical characteristic mass values

for a properly adjusted instrument are usually given in the instrument documen-tation.

Experimental values for characteristic mass (mo)can be determined for

comparison by measuring the peak area absorbance of a known mass of analyte

and calculating mo according to the following equation.

mo (pg)=Sample vol.(mL)x Analyte Conc.(mg/L)x 0.0044 AÃ s

Observed Peak Area (A×s)

Note that for this equation to be valid,the analyte mass used to measure mo must

produce a signal in the linear range of the calibration curve.By comparing the ex-perimental

value obtained from this equation to the reference value given in the

instrument documentation,the state of instrument optimization can be evaluated.

If the calculated mo is significantly greater than the reference value,adjustments

should be made to improve the sensitivity of the measurement.Calculated values.for mo which are significantly less than the reference value,suggesting better than

specified sensitivity,may in fact be a warning sign of analyte contamination in the

standard.

Another term characterizing graphite furnace instrument performance is ‘‘detec-tion

limit’’.Similar to flame AA,the detection limit is an indicator of the lower

limits of analyte detectability and is limited by the instrument signal-to-noise ratio.

Graphite furnace detection limits are usually stated in mass units (pg)rather than

concentration units,again reflecting the furnace signal’s dependency on analyte

mass,rather than concentration.However,in real analytical situations the detec-tion

limit will depend both on the system sensitivity (characteristic mass)and the

maximum sample size which can be accommodated.

Graphite Furnace Applications

The sensitivity of graphite furnace atomic absorption makes it the obvious choice

for trace metal analysis applications.Routine determinations at the mg/L level for

most elements make it ideal for environmental applications.Advances in instru-mentation

and techniques have made it possible to analyze very complex sample

matrices,such as those frequently found in biological and geological samples.

The microliter sample sizes used offer additional benefits where the amount of

sample available for analysis is limited,as in many clinical analyses.

COMPONENTS OF THE GRAPHITE FURNACE SYSTEM

The graphite furnace is made up of three major components,the atomizer,the

power supply,and the programmer.The atomizer is located in the sampling com-partment

of the atomic absorption spectrometer,where sample atomization and

light absorption occur.The power supply controls power and gas flows to the at-omizer

under the direction of the programmer,which is usually built into the

power supply or spectrometer.A description of each of these major components

follows.

The Graphite Furnace Atomizer

A basic graphite furnace atomizer is comprised of the following components:

· graphite tube

· electrical contacts

· enclosed water cooled housing

· inert purge gas controls

A graphite tube is normally the heating element of the graphite furnace.The cy-lindrical

tube is aligned horizontally in the optical path of the spectrometer and

serves as the spectrometer sampling cell.A few microliters (usually 5-50)of sam-ple

are measured and dispensed through a hole in the center of the tube wall onto

the inner tube wall or a graphite platform.The tube is held in place between two

graphite contact cylinders,which provide electrical connection.An electrical po-tential

applied to the contacts causes current to flow through the tube,the effect

of which is heating of the tube and the sample.

The entire assembly is mounted within an enclosed,water-cooled housing.Quartz

windows at each end of the housing allow light to pass through the tube.The

heated graphite is protected from air oxidation by the end windows and two

streams of argon.An external gas flow surrounds the outside of the tube,and a

separately controllable internal gas flow purges the inside of the tube.The system

should regulate the internal gas flow so that the internal flow is reduced or,pref-erably,

completely interrupted during atomization.This helps to maximize sample

residence time in the tube and increase the measurement signal.Figure 5-1 illus-trates

one type of atomizer assembly,a longitudinally-heated furnace.

The tube in Figure 5-1 is heated by passing electrical current from the graphite

contacts at the ends of the tube through the length of the tube.This type of furnace

is similar to the original design of Massmann,which is the basis for most currently

available commercial graphite furnace systems.

The longitudinally-heated furnace has a major liability.The electrical contacts at

each end of the tube must be cooled.As a result,there must always be a tempera-ture

gradient along the length of the tube,the tube ends adjacent to the electrical

contacts being cooler than the central portion.This temperature gradient can cause

vaporized atoms and molecules to condense as they diffuse to the cooler tube ends.

This may produce interferences,the most common type being the incomplete re-moval

of analyte or matrix from the tube.Incomplete removal of matrix during

pyrolysis can increase the magnitude of background absorption during atomiza-tion.

Incomplete removal of analyte during atomization is more serious.It creates

"carryover"or "memory",wherein a portion of the analyte in the current sample

remains in the tube and contributes to the analytical signal for the following sam-ple.

This produces erroneously high analytical results and poor precision.

To minimize carryover,most longitudinally-heated furnace heating programs use

one or more cleanout steps after the atomization step.A cleanout step involves the

application for several seconds of full internal gas flow and a temperature equal

to or greater than that used for atomization to remove residual sample components.

While this technique works well for the more easily atomized analytes,it is not

always successful with those analytes that require higher atomization tempera-tures.

The use of a high temperature cleanout step may also reduce tube lifetime.

The transversely-heated graphite furnace eliminates many of the problems asso-ciated

with the longitudinally-heated furnace.The graphite tube of the trans-versely-

heated furnace,shown in Figure 5-2,includes integral tabs which protrude

from each side.These tabs are inserted into the electrical contacts.When power

is applied,the tube is heated across its circumference (transversely).By applying

power in this manner,the tube is heated evenly over its entire length,eliminating

or significantly reducing the sample condensation problems seen with longitudi-nally-

heated furnace systems.

An additional advantage of the transversely-heated furnace is that it allows the use

of longitudinal Zeeman-effect background correction.As described in Chapter 3,

longitudinal Zeeman offers all of the advantages of transverse Zeeman correction

without the need to include a polarizer in the optical system.This provides a sig-nificant

improvement in light throughput.

The Graphite Furnace Power Supply and Programmer

The power supply and programmer perform the following functions:

· electrical power control

· temperature program control

· gas flow control

· spectrometer function control

The power supply controls the electrical current supplied to the graphite tube,

which causes heating.The temperature of the tube is controlled by a user-specified

temperature program.Through the programmer the operator will enter a sequence

of selected temperatures vs.time to carefully dry,pyrolyze,and finally atomize

the sample.The program may also include settings for the internal inert gas flow

rate and,in some cases,the selection of an alternate gas.Certain spectrometer

functions,such as triggering of the spectrometer read function,also may be pro-grammed

and synchronized with the atomization of the sample in the furnace.

SUMMARY OF A GRAPHITE FURNACE ANALYSIS

A graphite furnace analysis consists of measuring and dispensing a known volume

of sample into the furnace.The sample is then subjected to a multi-step tempera-ture

program.When the temperature is increased to the point where sample atomization occurs,the atomic absorption measurement is made.Variables under op-erator

control include the volume of sample placed into the furnace and heating

parameters for each step.These parameters include:

1)temperature final temperature during step

2)ramp time time for temperature increase

3)hold time time for maintaining final temperature

4)internal gas gas type and flow rate

In addition to the above,spectrometer

control functions can be programmed

to occur at specified times within the

graphite furnace program.While the

number of steps within each program

is variable,6 steps make up the typical

graphite furnace program.These

steps include:

1)Drying

2)Pyrolysis

3)Cool down (optional)

4)Atomization

5)Clean out

6)Cool down

Figure 5-3 illustrates a typical graph-ite

furnace program.The following

paragraphs will discuss each operator

controlled variable,and how they

may affect the analysis.

Sample Size

Since the graphite furnace signal depends on analyte mass,the operator has an ef-fective

degree of control on measured absorbance by controlling the sample vol-ume.

Larger volumes of sample solution contain more analyte and result in greater

signals.The analytical range of furnace analysis can therefore be controlled,to

some extent,by varying sample volume.For very low concentrations,the maximum volume of analyte can be used,while for higher concentrations,the sample

volume can be reduced.Smaller sample volumes can also be used where sample

availability is limited or where background absorption is excessively large.

The maximum volume of sample usable will depend on the tube configuration.

Where the graphite platform is not used,sample volumes up to 100 mL can be used,

depending on the type of tube and sample.With the platform in place,a sample

volume of less than 50 mL is recommended.A convenient sample volume for most

analyses is 20 mL.Where larger volumes are required,i.e.,for improved detection

limits,multiple injections can be used with appropriate drying and pyrolysis steps

between each injection to increase the effective sample size.

The use of an autosampler is strongly recommended for dispensing samples into

a graphite furnace.While skilled operators may obtain reasonable reproducibility

by manual injection on a short term basis,autosamplers have been proven to pro-vide

superior results.With many graphite furnace systems,autosamplers can also

generate working standards from stock standard solutions;add appropriate re-agents;

and provide method of additions analyses or recovery measurements,all

automatically.

The Drying Step

After the sample is placed in the furnace,it must be dried at a sufficiently low tem-perature

to avoid sample spattering,which would result in poor analytical preci-sion.

Temperatures around 100-120 o C are common for aqueous solutions.

Use of a temperature ‘‘ramp’’provides a variable time over which the temperature

is increased.A longer ramp time provides a slower,more ‘‘gentle’’increase in

heating.When a platform is used,the temperature lag of the platform versus the

tube walls provides a natural ‘‘ramping’’effect.Therefore shorter ramp times are

usually used with the platform.Longer ramp times are used when the sample is

to be atomized from the tube wall.

After the temperature ramp,the furnace is held at the selected drying temperature

until drying is complete.Since only a few microliters of sample are used,the dry-ing

‘‘hold’’time is usually less than a minute.

During the drying process,the internal gas flow normally is left at its default maxi-mum

value (250-300 mL per minute)to purge the vaporized solvent from the tube.

The Pyrolysis Step

The purpose of the pyrolysis step (sometimes referred to as the ashing,char or pre-treatment

step)is to volatilize inorganic and organic matrix components selec-tively

from the sample,leaving the analyte element in a less complex matrix for

analysis.During this step,the temperature is increased as high as possible to vola-tilize

matrix components but below the temperature at which analyte loss would

occur.

The temperature selected for the pyrolysis step will depend on the analyte and the

matrix.Suggested temperatures normally are provided in the documentation sup-plied

with the graphite furnace.The internal gas flow is again left at 250-300 mL

per minute in the pyrolysis step,to drive off volatilized matrix materials.For some

sample types,it may be advantageous to change the internal gas,e.g.,to air or oxy-gen,

during the pyrolysis step to aid in the sample decomposition.

The Pre-atomization Cool Down Step

With longitudinally-heated furnaces,it is frequently advantageous to cool the fur-nace

prior to atomization since the heating rate is a function of the temperature

range to be covered.As the temperature range is increased,the rate of heating also

increases.The use of a cool down step prior to atomization maximizes the heating

rate and extends the isothermal zone within the tube immediately after heating.

The extended isothermal zone has been shown to improve sensitivity and reduce

peak tailing for a number of elements,including those which characteristically are

difficult to atomize in the graphite furnace.

A pre-atomization cool down step normally is not required for transversely-heated

furnaces as the isothermal zone extends the length of the tube with that type of

system.

The Atomization Step

The purpose of the atomization step is to produce an atomic vapor of the analyte

elements,thereby allowing atomic absorption to be measured.The temperature in

this step is increased to the point where dissociation of volatilized molecular spe-cies

occurs.

The atomization temperature is a property of the analyte element.By following

recommended procedures of analysis,it is usually possible to use the temperatures

provided in the graphite furnace documentation without further optimization.

Care should be taken to avoid the use of an excessively high atomization tempera-

ture,as the analyte residence time in the tube will be decreased and a loss of sen-sitivity

will occur.Also,the use of excessively high atomization temperatures can

shorten the useful lifetime of the graphite tube.

For atomization,it is desirable to increase the temperature as quickly as possible.

Therefore,ramp times normally will be set to minimum values to provide the high-est

heating rate.It also is desirable to reduce or,preferably,to totally interrupt the

internal gas flow during atomization.This increases the residence time of the

atomic vapor in the furnace,maximizing sensitivity and reducing some interfer-ence

effects.At the beginning of this step,the spectrometer ‘‘read’’function is trig-gered

to begin the measurement of light absorption.

The Clean Out and Cool Down Steps

After atomization,the graphite furnace may be heated to still higher temperatures

to burn off any sample residue which may remain in the furnace.An optional cool

down step then allows the furnace to return to near ambient temperature prior to

the introduction of the next sample.With some systems,a preset cool down step

is automatically included in each furnace cycle,and need not be programmed sepa-rately.

FAST FURNACE ANALYSIS

The most time consuming portions of a graphite furnace AA analysis are the dry-ing

and pyrolysis steps.Analysts have long sought some means to reduce or elimi-nate

the time required for these pretreatment stages.For years,an inability to

compensate for the high background signals generated with incomplete pretreat-ment

and matrix-related interferences precluded reducing the time required for

these steps.

The situation changed dramatically with the introduction of two developments:

Zeeman effect background correction and Stabilized Platform Graphite Furnace

(STPF)technology,which is described in detail in Chapter 6.Zeeman effect back-ground

correction is far superior to continuum source background correction in

its ability to correct accurately for high levels of background absorption.The use

of STPF technology provides almost constant characteristic mass values inde-pendent

of sample matrix.The combination of these two techniques was the key

to providing faster graphite furnace analyses.

In fast furnace analysis,the pyrolysis step and matrix modification are usually

eliminated.The drying step is minimized by injection of the sample onto a pre-heated

platform.Drying time is also minimized by using as small a sample size

as possible consistent with the required analytical precision and detection limits.

To further reduce the furnace program time,the cool down step at the end of the

program is frequently reduced or eliminated.This is possible since most furnace

systems typically require 20 seconds or more between the end of a temperature

program and the point at which the next sample is ready to be introduced into the

furnace.That 20 second period is usually sufficient to allow the furnace to cool

reproducibly to the preheated drying temperature required for the next sample.

Using these procedures,typical furnace program times frequently can be reduced

from 2-4 minutes to 30 seconds or less per determination,a considerable time sav-ings,

without sacrificing analytical precision or accuracy.

Fast furnace analysis techniques are not compatible with all AA instrumentation.

To fully realize the benefits of fast furnace analysis,the instrumentation used must

provide Zeeman background correction,be capable of handling very high absor-bance

measurements and be compatible with the requirements of STPF technol-ogy.

Also,the fast furnace technique may not be compatible with all sample types.

Very complex matrices may still require at least a short pyrolysis step and use of

a matrix modifier for optimum results.

MEASURING THE GRAPHITE FURNACE AA SIGNAL

Nature of the Graphite Furnace Signal

In flame atomic absorption,the absorption signal is steady state.That is,as long

as solution is aspirated into the flame,a constant absorbance is observed.For

graphite furnace analyses,however,the signal is transient.As atomization begins,

analyte atoms are formed and the signal increases,reflecting the increasing atom

population in the furnace.The signal will continue to increase until the rate of atom

generation becomes less than the rate of atom diffusion out of the furnace.At that

point,the falling atom population results in a signal which decreases until all at-oms

are lost and the signal has fallen to zero.To determine the analyte content of

the sample,the resulting peak-shaped signal must be quantitated.

Peak Height Measurement

For many years,measuring the height of the transient signal was the only practical

means for quantitating furnace results.The constantly changing signal was moni-tored

on a strip chart recorder,and peak height was measured manually in chart

divisions.Later instrumentation allowed peak height to be measured directly by

electronic means.

While peak height does depend on the analyte concentration in the sample,it is

also affected by other factors.Peak height is only a measure of the maximum atom

population which occurred in the furnace during atomization.If matrix compo-nents

in the sample affect the rate of atom formation,the maximum atom popu-lation

and the peak height are also affected,as shown in Figure 5-4 for the

determination of lead in blood.While the two solutions contain identical amounts

of lead (0.2 ng),the peak shapes and appearance times are dramatically different.

This susceptibility to matrix effects makes graphite furnace AA vulnerable to in-terferences

when peak height measurement is used for quantitation.Therefore,

peak height measurements are seldom used with modern graphite furnace AA sys-tems.

Peak Area Measurement

Modern instrumentation provides the capability to integrate absorbance during the

entire atomization period,yielding a signal equal to the integrated peak area,that

is,the area under the peak signal.If the temperature in the furnace is constant dur-ing

the measurement process,the peak area will represent a count of all atoms pre-sent

in the sample aliquot,regardless of whether the atoms were generated early

or late in the atomization process.Integrated peak area measurements (A .s)are

independent of the atomization rate,and are therefore much less subject to matrix

effects as shown in Figure 5-4.As a result,peak area is preferred for graphite fur-nace

analysis.